Regulatory update
EU MDR 2017/745 FDA

ISO 10993-1:2025 is here. What changed, and how to bring your documentation in line.

The five structural changes in Edition 6, explained from the perspective of a Technical Documentation reviewer.

About the author

James Ward, PhD

James is Klaris Regulatory Lead. He was previously Principal Technical Specialist and Decision Maker (Panel Member) at BSI, focused on biological evaluation and chemistry, where he reviewed Biological Safety documentation across a wide range of medical devices. He now leads the requirements library that powers Klaris Biological Evaluation Review and Remediation.

View on LinkedIn ↗

ISO 10993-1:2025 (Edition 6) is the new benchmark for the biological evaluation of medical devices. It brings five structural changes, and it is the version Notified Bodies now read as the state of the art under EU MDR, recognized in part by the FDA.

The edition anchors biological evaluation inside the ISO 14971:2019 risk process, widens what the exposure profile has to cover, and raises expectations for chemical characterization and animal welfare. Because the FDA has recognized Edition 6, US manufacturers have until July 1, 2029 to migrate.

This article walks through all five changes, one clause at a time. For each, you get a reviewer's read of what a Notified Body or FDA reviewer is looking for, then the smallest practical move that closes the gap. It is written by James Ward, PhD, Klaris's regulatory lead, drawing on years spent reviewing technical documentation inside a Notified Body. A printable migration checklist and a jurisdiction-by-jurisdiction guide are at the end.

Timeline: ISO 10993-1:2025, EU MDR, and FDA.

  1. 2009 ISO · Edition 4 Risk-based footing. First explicit framing as a risk-management activity.
  2. 2018 ISO · Edition 5 Categorization as prerequisites for consideration, not a checklist of tests.
  3. 2021 EU MDR · Date of application EU MDR 2017/745 applies. Biological safety must be evaluated inside the ISO 14971:2019 risk process.
  4. 2025 ISO · Edition 6 Anchored in ISO 14971:2019. Four contact-tissue tables. The edition Notified Bodies increasingly read as state of the art under EU MDR. You are here
  5. 2029 FDA · Transition FDA transition closes 1 July 2029. After it, only declarations to the 2025 edition are accepted; the 2018 edition (RN 2-258) is retired. Upcoming

ISO 10993-1:2025 key changes

Five structural changes, each its own section below. Pick a row to jump straight to it.

No. Change Under 2018 Under 2025
Categorization One Table A.1 for every device Four contact-tissue tables under § 6.4; duration thresholds shifted
Foreseeable misuse The IFU bounds the exposure profile Reasonably foreseeable misuse is in scope
Risk anchoring Biological testing as a standalone checklist Every biological effect traces to a hazard under ISO 14971:2019
Chemical characterization One option among several The default starting point; E&L and toxicological assessment up front
Lifecycle A deliverable, frozen at approval A lifecycle activity; re-evaluation on every trigger
ISO 10993-1:2025 § 6.4 FDA 2023 guidance, Attachment A

Recategorization, one table became four.

The single Table A.1 from 2018 is now four contact-tissue tables. Many devices move category.

The 2018 edition had a single categorization table. § 6.4 of the 2025 edition reorganizes categorization across four contact-tissue tables, with contact-duration thresholds adjusted in several places. For reusable devices, total contact is now read cumulatively: repeated short uses can add up to a longer-duration bracket even when no single use is long.

A device categorized correctly under 2018 may not be categorized correctly under 2025. The test panel that follows from the category drives the entire BER. A miscategorization under the new tables means the panel itself is wrong, which is a structural gap, not a paperwork one.

2018
Table A.1
One table for everything.
2025 · § 6.4
Table 1
Table 2
Table 3
Table 4
Four contact-tissue tables. Contact-duration thresholds shifted.
Devices near a category boundary, especially circulating blood and contact close to 24 h or 30 d, are the ones most likely to move.

Spot it

Open the BEP. If it cites the 2018 Table A.1, the categorization is stale by definition. Pay closest attention to devices near a boundary, circulating-blood contact or duration close to 24 h or 30 d, which are the ones most likely to cross into a new bracket.

Fix it

Recategorize against the four § 6.4 tables before rewriting anything else, and update the BEP to cite the 2025 edition. Starting the rewrite from the old panel is the common way to lose weeks of editing. Where the panel grows, plan the new biological effects before commissioning tests: an unjustified gap is easier to defend in remediation than a partial test that hints at a missing effect.

Klaris Biological Evaluation Review and Remediation Klaris flags any device whose 2018 category no longer fits the 2025 four-table framework. Learn more
ISO 10993-1:2025 § 4.2, § 5.2

Foreseeable misuse, in scope.

The IFU no longer fully bounds the exposure profile.

The 2025 edition is explicit that biological risks must be assessed across both intended use and reasonably foreseeable misuse. Extended contact duration, improper reprocessing, use in an unintended biological environment, use in unintended patient populations: these are all now in scope. Abstract scenarios unlikely to occur are not. The IFU is one input to the exposure profile, not the boundary of it.

Two patterns surface in review: a device used past the duration the IFU restricts it to, and a device used at a contact site the IFU does not name. Both can move you from "limited exposure" to "long-term", which opens up biological effects (chronic toxicity, local effects after tissue contact) that were not in the original panel.

Spot it

Look at how the BEP derives its exposure category. If it comes straight from the IFU with no consideration of off-label or foreseeable-misuse exposure, the file is incomplete by the new standard.

Fix it

Add a short section to the BEP naming the foreseeable-misuse cases considered, the exposure each would imply, and either the extra biological effects run to cover them or the argument that the original panel already does. The slow part is usually getting the clinical and regulatory teams to agree on the list, not the edit itself; post-market complaints and competitor IFUs are a good way to seed it. A covered scenario or a reasoned exclusion is acceptable; silence is not.

Klaris Biological Evaluation Review and Remediation Klaris cross-checks the IFU against the BEP and surfaces inconsistencies in the exposure and foreseeable-misuse coverage. Learn more
ISO 10993-1:2025 § 4, § 6 · ISO 14971:2019 21 CFR 820.30(g)FDA RN 2-313 (§ 6.9 carve-out)

Anchored in ISO 14971:2019.

Every biological effect must trace back to a hazard in your risk file.

Edition 6 re-frames biological evaluation as a risk-management activity. Every biological effect should be traceable to an identified hazard in your risk file, and every residual risk should be expressed in the vocabulary the rest of your risk documentation uses. The standard is explicit that biological evaluation is not synonymous with biological testing: testing follows from the risk analysis, not the other way around.

When a Notified Body picks up your BER, the first question is no longer "did you run the right tests" but "did you run the right tests for the hazards you identified". A test panel that does not map to your hazard register reads as evidence of a disconnect between biological evaluation and the rest of the technical documentation.

Spot it

Open the BER, pick three biological effects, and for each find the matching hazard ID in the risk-management file. If you cannot, the file fails the new test. Many 2018-era BERs leave that mapping implicit, which is exactly what no longer passes.

Fix it

Add a single hazard-to-effect cross-reference table to the BER, and update the risk file so each biological hazard has a control entry pointing at the relevant 10993-series report. The trace usually already exists, held implicitly across the risk file, BER, and supporting reports and in the team's heads; pulling it into one table is a week of careful work for whoever knows the file best, not an evening. Resist inventing retroactive hazards to fill the trace: a test run with no hazard behind it is itself a gap to discuss.

Klaris Biological Evaluation Review and Remediation Klaris surfaces inconsistencies between the BER's biological effects and the hazards, warnings, precautions, and residual risks stated in the risk file and IFU. Learn more
ISO 10993-18:2020 · ISO 10993-17:2023 FDA recognized consensus standardsFDA 2023 guidance, Attachment A

Chemical characterization, the default starting point.

ISO 10993-18:2020 work is now the default starting point. Absence needs a written defense.

The 2025 edition elevates chemical characterization from one option among several to the default starting point for most biological evaluations. The toxicological risk assessment that follows from it (per ISO 10993-17:2023) is now expected up front. Where the 2018 edition allowed broad justification for skipping these, the 2025 edition expects either the studies themselves or a defensible written argument for their absence, anchored in the device's risk class and exposure profile.

"Material composition is documented" is no longer enough on its own for higher-risk devices: the reviewer expects either an E&L study with a toxicological assessment, or a written rationale for its absence that holds up against the device's risk class and exposure profile. Genotoxicity in particular has widened, with the 2025 edition expecting a documented rationale rather than a default skip, so pair ISO 10993-3:2014 with the toxicological assessment under ISO 10993-17:2023.

Spot it

Search the BER for "extractables", "leachables", "ISO 10993-18", or "ISO 10993-17". If they are absent on a Class IIb or III device with anything beyond limited exposure, the file is likely short, and a missing genotoxicity rationale on any prolonged- or permanent-contact device is the same kind of gap. Equivalence arguments that lean on the absence of chemical characterization are especially exposed.

Fix it

For long-term and implant devices, plan an E&L study under exhaustive extraction conditions paired with a toxicological risk assessment per ISO 10993-17:2023, and book it early: the bottleneck is rarely the study but getting the toxicologist and sample stock lined up at once, which is months, not weeks. For lower-risk devices, write the rationale for not running E&L into the BEP, referencing the categorization that drives the decision. Either path is defensible; neither being documented is not.

Klaris Biological Evaluation Review and Remediation Klaris checks the BER for E&L coverage and the BEP for the written rationale. Learn more
ISO 10993-1:2025 § 4.6, § 6.6.2, § 7

Lifecycle, not deliverable.

Design, supplier, material, and post-market signals all trigger reassessment.

The 2025 edition treats biological evaluation as an activity that runs across the device lifecycle. Design changes, material changes, supplier changes, manufacturing changes, complaints, and post-market signals all trigger reassessment. § 6.6.2 sets the expectation for existing devices specifically: files requiring assessment of additional biological effects need to be updated in the short term; others may not need an update yet, provided the justification is documented.

A frozen BER attached to a CE mark from 2021, with no review since, is no longer acceptable on its own. The reviewer expects a documented review cadence and clear evidence of reassessment when relevant signals occurred.

Spot it

Compare the BER's revision history against the device's change log. If the last biological-evaluation update predates the most recent design, supplier, or material change, the file is not lifecycle-current.

Fix it

Define a written re-evaluation trigger list, design, supplier, material, manufacturing, complaint, and post-market signal, and add a short re-evaluation log to the BER that records each trigger and its outcome, even when the outcome is "no change required". Wire the list into the QMS change-control SOP rather than leaving it standalone: if the SOP forces a biocomp checkbox on every design or supplier change, the log fills itself; if it does not, it quietly falls behind. The log is the artifact the reviewer looks for.

Klaris Biological Evaluation Review and Remediation Klaris checks your re-evaluation record against the change history and the standard's trigger requirements, and flags what's missing. Learn more
Source: https://www.klaris.ai/insights/iso-10993-1-2025-transition · Published 3 Jun 2026 · Last reviewed 3 Jun 2026
EU MDR vs US FDA adoption

How ISO 10993-1:2025 lands across jurisdictions.

EU MDR State of the art (not harmonized)

ISO 10993-1:2025 is not harmonized under EU MDR, so it grants no automatic presumption of conformity. Notified Bodies still treat it as the current state of the art for biocompatibility and risk-based safety evaluation, so your biological evaluation plan and report should follow it to meet their expectations.

FDA Recognized in the CDRH database (partial)

FDA recognizes the 2025 sixth edition in its CDRH consensus-standards database (RN 2-313, partial), with specific clauses carved out (risk estimation defers to ISO 14971:2019); declarations to the 2018 fifth edition stay acceptable until 1 July 2029. FDA also applies its own endpoint matrix through Attachment A of the 2023 guidance.

FDA recognition, in detail

Risk estimation
Clause 6.9 defers to clauses 5.5, 6, and 7 of ISO 14971:2019.
Scope carve-out
The phrase “consumer products or” in clause 6.5.11.3 is excluded; it conflicts with Attachment G of the 2023 guidance.
Transition
Declarations to the 2018 fifth edition stay acceptable until 1 July 2029.
Genotoxicity
The added requirements in Tables 2–4 and clause 6.5.7 may not line up with Table A.1 in Attachment A for prolonged-contact devices; settle that battery in a Pre-Submission.
Terminology
FDA keeps “endpoint matrix” where the 2025 edition says “biological effect”, the same panel under two labels.
Where Klaris fits

Catch the 2025 gaps
before your reviewer does.

The judgment is yours; the gap-finding is ours. Klaris reads your biological safety documentation, risk file, and IFU against ISO 10993-1:2025 overnight, and hands back every gap and inconsistency with its clause already cited, ready for your Notified Body and your FDA reviewer.

A migration checklist.

Print this, run it against any 2018-era documentation before you resubmit, then attach the result to your QMS.

10 checks

Notes. The “Reviewer opinion” notes reflect the author's review experience on both sides of the file and are not endorsements by any Notified Body or regulator.

Disclosures. James Ward is Klaris's Regulatory Lead.

Sources.

Stay in the loop

New regulatory insights, straight to your inbox.

Stay up to date and remediate before your reviewer asks.

No spam, just new regulatory insights. Unsubscribe anytime. By subscribing you agree to our Privacy Policy.